Phthalates are esters of phthalic acid that have extensively been used as plasticizers to increase flexibility, transparency, durability, and longevity in a wide variety of consumer and household products, such as children’s toys, electronics, clothes, flooring, wallpaper, and paints. Phthalates are also used, as plasticizers, solubilizers, or denaturants in cosmetics and personal care products, such as perfumes, nail polishes, and hair sprays. Many phthalates are classified as hazardous because of their effects on the reproductive system and their association with an increased risk of cancer.
Parabens are esters of parahydroxybenzoic acid, which due to their low volatility, high stability, antibacterial and antifungal properties, have been used as preservatives in cosmetics, personal care, pharmaceutical, food, and industrial products. Parabens have been associated with allergenic contact dermatitis and rosecea.
Triclocarban is an antibacterial and antifungal agent that is used in many cosmetic and personal care products, including soap, toothpaste, deodorant, shampoo and shaving cream. Triclocarban is also used in several consumer products including kitchen cutting boards, shoes, towels, and clothing, as well as in medical disinfectants and medical products. But there are several health concerns related to the use of triclocarban, including potential hormone and endocrine disruption, as well as its potential to contribute to the development of antibiotic resistance.
In order to accommodate consumer demands for higher standards, many manufacturers are developing, and labeling cosmetic and personal care products ‘free from’ phthalates and parabens.
In this application note, we describe the method development, sample extraction, and mass spectral analysis of parabens, phthalates, and triclocarban using the ACQUITY UPLC H-Class System, coupled to the ACQUITY QDa Detector. Learn how mass detection can be used to achieve reliable analytical methods to unequivocally identify and quantify compounds such as phthalates, parabens, and triclocarban, during both method development stages, and during routine regulatory analysis.
The ACQUITY QDa Detector linked to the ACQUITY UPLC H-Class System provides improved confidence in the identification and quantification of phthalates and parabens in cosmetics and personal care products offering:
Phthalates are esters of phthalic acid that have extensively been used as plasticizers to increase flexibility, transparency, durability, and longevity in a wide variety of consumer and household products, such as children’s toys, electronics, clothes, flooring, wallpaper, and paints. Phthalates are also used, as plasticizers, solubilizers, or denaturants in cosmetics and personal care products, such as perfumes, nail polishes, and hair sprays.
Parabens are esters of parahydroxybenzoic acid, which due to their low volatility, high stability, antibacterial and antifungal properties, have been used as preservatives in cosmetics, personal care, pharmaceutical, food, and industrial products.
Triclocarban is an antibacterial and antifungal agent that is used in many cosmetic and personal care products, including soap, toothpaste, deodorant, shampoo and shaving cream. Triclocarban is also used in several consumer products including kitchen cutting boards, shoes, towels, and clothing, as well as in medical disinfectants and medical products. But there are several health concerns related to the use of triclocarban, including potential hormone and endocrine disruption, and also its potential to contribute to the development of antibiotic resistance.
Many phthalates are classified as hazardous because of their effects on the reproductive system and their association with an increased risk of cancer. Parabens are associated with allergenic contact dermatitis and rosecea. Studies1,2 have also suggested parabens may be carcinogenic and possess estrogenic disrupting activities. Due to these properties phthalates, parabens, and triclocarban are either banned or restricted, as regulated by the Cosmetic Directive 1223/2009.3
In order to accommodate consumer demands for higher standards, many manufacturers are developing, and labeling cosmetic and personal care products ‘free from’ phthalates and parabens.
Previous example methodologies for the analysis of phthalates include GC-MS,4 and HPLC-UV4; GC-FID,5 HPLC-UV,4,6 HPLC-MS,7 GC-MS,4 and capillary electrophoresis6 for the analysis of parabens; and HPLC-MS8 for the analysis of triclocarban.
Accessible and intuitive as an optical detector, the ACQUITY QDa Detector has been designed for chromatographers with ease of use in mind. Mass detection can be used to achieve reliable analytical methods to unequivocally identify and quantify compounds such as phthalates, parabens, and triclocarban, during both method development stages, and during routine regulatory analysis.
This application note considers the method development, sample extraction, and mass spectral analysis of parabens, phthalates, and triclocarban using Waters ACQUITY UPLC H-Class System, coupled to the ACQUITY QDa Detector.
LC system: |
ACQUITY UPLC H-Class |
Runtime: |
5.00 min |
Column: |
ACQUITY UPLC BEH C18, 1.7 μm, 2.1 x 50 mm |
Column temp.: |
40 °C |
Sample temp.: |
10 °C |
Mobile phase A: |
Water + 0.1% formic acid |
Mobile phase B: |
Methanol + 0.1% formic acid |
Flow rate: |
0.6 mL/min |
Injection volume: |
5.0 μL |
|
Time (min) |
Flow rate (mL/min) |
%A |
%B |
Curve |
---|---|---|---|---|---|
1.0 |
Initial |
0.6 |
30.0 |
70.0 |
|
2.0 |
1.0 |
0.6 |
30.0 |
70.0 |
6.0 |
3.0 |
1.5 |
0.6 |
10.0 |
90.0 |
6.0 |
4.0 |
4.0 |
0.6 |
10.0 |
90.0 |
6.0 |
5.0 |
4.01 |
0.6 |
30.0 |
70.0 |
6.0 |
6.0 |
5.0 |
0.6 |
30.0 |
70.0 |
6.0 |
MS system: |
ACQUITY QDa |
Ionization mode: |
ESI + and - |
Capillary voltage: |
0.8 kV |
Probe temp.: |
450 °C |
Acquisition: |
Selected Ion Recording (SIR) |
Cone voltage: |
15 V |
The list of compounds considered, including phthalates, parabens, and triclocarban, along with their expected retention times are detailed in Table 2. The empirical formulas and structures are detailed in Tables 3 and 4.
Empower 3 Software was used to control the ACQUITY UPLC H-Class System and the AQCUITY QDa Detector, as well as for data acquisition and quantitation.
A fast, selective, and sensitive LC-MS method for the detection of a selection of phthalates, parabens, and triclocarban in cosmetic and personal care products has been developed.
The ACQUITY QDa Detector’s SIR parameters were optimized, considering both negative and positive electrospray ionization modes, in order to ensure full coverage of all the compounds being analyzed (as detailed in Table 2.)
Method development was carried out using reversed phase UPLC, where different gradient conditions, columns, and mobile phases were considered. The objective was to separate the isomeric phthalate compounds considered: di-n-octyl phthalate (DiNP), and diisobutyl phthalate (DiBP); bis(2-ethylhexyl) phthalate (DEHP), and di-n-octyl phthalate (DnOP) – while maintaining sample throughput. This was achieved by optimizing the mobile phases and the gradient eluting conditions used. The final LC conditions used are detailed in the methods section.
The method was established over the calibration ranges of 0.01 µg/mL to 10 µg/mL for phthalates and triclocarban, and 0.05 µg/mL to 25 µg/mL for parabens, equivalent to 0.25 to 250 mg/Kg, and 1.25 to 625 mg/Kg in the extracted samples respectively. Good linearity was achieved for all the compounds considered (R2 >0.99). SIR chromatograms for phthalates, parabens, and triclocarban in a mixed 1.0 µg/mL calibration standard are shown in Figure 1.
The developed five-minute UPLC method, is more than seven times faster than existing HPLC and GC methods, with an excess of 90% less solvent usage than existing HPLC methods.
The SIR mass detection conditions detailed in Table 2 were used after appropriate sample preparation to screen for phthalates, parabens, and triclocarban in cosmetic and personal care samples.
Samples were fortified at various levels with selected phthalates and parabens, then prepared for analysis as detailed in the experimental section. Example SIR chromatograms achieved are shown in Figure 2.
720005521, October 2015