The APC with QSM is specifically designed for aggressive solvents used in the polymer industry. The Waters ACQUITY Advanced Polymer Chromatography (APC) System is designed to operate consistently using harsh solvents like THF, chloroform, and dimethyl sulfoxide. By adding the flexibility of a quaternary pump to the industry-accepted APC system, polymer analysis techniques are no longer limited to HPLC technology.
The gradient polymer elution chromatography (GPEC) chromatogram in Figure 1 has a run time of 30 minutes using an Alliance HPLC (high-performance liquid chromatography) system. If this method is transferred to an UltraPerformance Liquid Chromatography (UPLC) system, the run time could be significantly shortened. However, most UPLC systems are not compatible with organic solvents, such as tetrahydrofuran (THF), and are not robust enough to constantly use 100% organic solvent.
The Waters ACQUITY Advanced Polymer Chromatography (APC) System is designed to operate consistently using harsh solvents like THF, chloroform, and dimethyl sulfoxide. Originally, the APC system was designed for gel permeation chromatography (GPC) using an isocratic pump. By adding the polymer quaternary solvent manager (p-QSM) for gradient elution and an evaporative light scattering detector (ELSD) to the APC system, GPEC can be accomplished in less than 10 minutes (Figure 1).
System: |
APC with p-QSM |
Column: |
XBridge BEH C8, XP, 2.5 μm, 3.0 × 100 mm (p/n 186006047) |
Mobile phase A/B: |
Methanol |
Mobile phase C/D: |
Tetrahydrofuran |
Flow rate: |
0.7 mL/min |
Column temp.: |
40 °C |
Injection volume: |
1 μL |
Sample concentration: |
PMMA (1 mg/mL), PS (0.12 mg/mL), PBD (0.5 mg/mL) |
Sr. No |
Time |
Flow (mL/min) |
%A |
%B |
%C |
%D |
Curve |
---|---|---|---|---|---|---|---|
1 |
Initial |
0.700 |
100.0 |
0.0 |
0.0 |
0.0 |
Initial |
2 |
1.00 |
0.700 |
60.0 |
0.0 |
40.0 |
0.0 |
11 |
3 |
2.00 |
0.700 |
35.0 |
0.0 |
65.0 |
0.0 |
11 |
4 |
3.00 |
0.700 |
15.0 |
0.0 |
85.0 |
0.0 |
11 |
5 |
4.00 |
0.700 |
15.0 |
0.0 |
85.0 |
0.0 |
11 |
6 |
5.00 |
0.700 |
0.0 |
0.0 |
100.0 |
0.0 |
6 |
7 |
6.00 |
0.700 |
100.0 |
0.0 |
0.0 |
0.0 |
6 |
8 |
7.00 |
0.700 |
100.0 |
0.0 |
0.0 |
0.0 |
11 |
In order to precipitate a polymer onto the column, the polymer solubility in the gradient blend of solvents needs to be determined. The accepted method in literature is to titrate a known concentration of polymer solution (10–2 to 10–3 mg/mL) with a non-solvent (poor solvent). As the titration progresses, the polymer solution will appear cloudy as the polymer aggregates and precipitates: the polymer is repelled by the non-solvent and the polymer molecules are forced together in aggregation. The volume fraction of the non-solvent is calculated using Equation 1, where phi (F) equals the cloud point (CP) divided by 100, which equals the volume of the non-solvent (Vns) divided by the sum of polymer volume, good solvent volume, and non-solvent volume (Vp + Vs + Vns). This technique is called cloud-point determination.4,5
Eq 1. F = CP/100 = Vns / Vp + Vs + VNS
If the linear gradient cannot be optimized to achieve reproducible peak areas, a step gradient may be needed. The GPEC of three polymers, polystyrene (PS), polybutadiene (PBD), and polymethylmethacrylate (PMMA), are used for this experiment (Table 1). The linear gradient produces poor peak area reproducibility for PS and PMMA, and the PBD does not elute.
Therefore, a step gradient method is used to control the amount of time the polymers spend in the insoluble regions of the gradient. A steep curve is used for the insoluble regions (11) of the gradient profile, while the soluble region is a less steep curve (6) as seen in the gradient table and resulting chromatography (Figure 2).
By using the Empower 3 CDS custom calculations and report features, there is no need to export data into a separate spreadsheet. The report in Figure 3 has the peak area ratios and the gradient ratios calculated within Empower. The complete GPEC project and detailed GPEC guide are available on the Waters website and in the Empower marketplace.
GPEC is not GPC. In the example chromatograms of Figure 4, three chemically different polymers co-elute using a GPC separation, because they have similar hydrodynamic volumes. If reversed-phase separation is used to separate these polymers, they would remain partially soluble throughout the separation and not baseline separate: partially overlapping. The GPEC method purposefully precipitates the polymers temporarily onto the column or column frit until the gradient ratio of good solvent reaches the polymer’s solubility point, releasing the polymer from the surface of the column. Instead of using the size-based separation of GPC, the polymers are separated by their chemical solubility using GPEC. GPEC is an LC application and is completely different from a GPC application.1
The APC system designed for GPEC has a new configuration (Figure 5). The polymer isocratic solvent manager (p-ISM) has been replaced with a polymer quaternary solvent manager (p-QSM) to accommodate gradient elution with 100% organic solvent. The refractive index (RI) detector has been replaced with an evaporative light scattering detector (ELSD), because RI is not compatible with gradient elution. A photodiode array (PDA) is not used for this application example, because not all polymers are UV active.2,3
GPEC is a complementary analysis to GPC, when co-eluting peaks interfere with determining polymer ratios in a mixture. This uncommon analysis technique has been used for decades but has been limited to traditional HPLC instruments. UPLC instruments have solvent compatibility limitations that can limit analysis options like GPEC. The APC with QSM is specifically designed for aggressive solvents used in the polymer industry. By adding the flexibility of a quaternary pump to the industry-accepted APC system, polymer analysis techniques are no longer limited to HPLC technology.
720006705 November 2019